Chapter 3, CDMA Receiver Tests
Measuring Receiver Sensitivity and Dynamic Range
Cha
p
te
r 3
C
D
M
A
R
ece
iver
Test
s
Measuring Receiver Sensitivity and Dynamic Range
The Test Set performs "Receiver Sensitivity and Dynamic Range" as described in
TIA IS-98, "CDMA Receiver Minimum Standards."
During this test, FER is measured with the Test Set first providing a low level
signal, then a high level signal.
NOTE:
The following procedure assumes that an instrument PRESET will be performed as part of
setting up a call. This clears all settings from other tests that may affect the accuracy of this
test, and allows the procedure to begin from a known instrument state.