Chapter 4, CDMA Transmitter Tests
Measuring Maximum RF Output Power
Measuring Maximum RF Output Power
The Test Set measures mobile-station-under-test (MSUT) power using the
average power meter. This test is performed according to TIA/EIA IS-98.
Test Prerequisites
•
The Average Power measurement must be zeroed. Zeroing the average power
measurement should be performed at least as often as the following conditions arise:
•
5 °C change in operating temperature
•
Power cycle
•
Daily
See "Zeroing Average Power Measurements" on page 36
return to this procedure.
NOTE:
The following procedure assumes that an instrument PRESET will be performed as part of
setting up a call. This clears all settings from other tests that may affect the accuracy of this
test, and allows the procedure to begin from a known instrument state.