Power Meter Remote Operation
Using Sensor Calibration Tables
1-30
HP E4418B/E4419B Programming Guide
Using Sensor Calibration Tables
This section applies to all HP 8480 series power sensors. It does not apply
to the HP E-series power sensors. The HP E-series power sensors have
their sensor calibration tables stored in EEPROM which allows frequency
and calibration factor data to be downloaded by the power meter
automatically.
This section describes how to use sensor calibration tables. Sensor
calibration tables are used to store the measurement calibration factors,
supplied with each power sensor, in the power meter. These calibration
factors are used to correct measurement results.
Overview
For the HP 8480 series power sensors there are two methods of providing
correction data to the power meter depending on the setting of the
[SENSe[1]]|SENSe2:CORRection:CSET1:STATe
command. If
[SENSe[1]]|SENSe2:CORRection:CSET1:STATe
is
OFF
the sensor
calibration tables are not used. To make a calibrated power measurement
when
[SENSe[1]]|SENSe2:CORRection:CSET1:STATe
is
OFF
, perform
the following steps:
1.
Zero and calibrate the power meter. Before carrying out the
calibration set the reference calibration factor for the power meter
you are using.
2.
Set the calibration factor to the value for the frequency of the
signal you want to measure.
3.
Make the measurement.
When
[SENSe[1]]|SENSe2:CORRection:CSET1:STATe
is
ON
, the
sensor calibration tables are used, providing you with a quick and
convenient method for making power measurements at a range of
frequencies using one or more power sensors. Note that with the sensor
calibration table selected, the RCF from the table overides any value
previously set. The power meter is capable of storing 20 sensor calibration
tables of 80 frequency points each.
Summary of Contents for E4418B
Page 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Page 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Page 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Page 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Page 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Page 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Page 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Page 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Page 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Page 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Page 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Page 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Page 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Page 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Page 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Page 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...