CALCulate Subsystem
CALCulate[1|2]:RELative[:MAGNitude]:AUTO <Boolean>|ONCE
3-26
HP E4418B/E4419B Programming Guide
CALCulate[1|2]:RELative[:MAGNitude]:AUTO <Boolean>|ONCE
This command is used to set the reference value to be used in the relative
measurement. Within the
CALCulate
block the relative value is applied
to the measurement signal after any math calculations and display offsets
have been applied.
This command should be set to
ONCE
to set the reference value to be used
in relative measurements. Selecting
ONCE
sets the reference value to
being that of the measurement signal after any math calculations and
display offsets have been applied. After the reference value has been set
the command returns to
OFF
. Setting this command to
ONCE
turns the
CALCulate[1|2]:RELative:STATe
command to
ON
.
If
0|OFF
is selected, no reference value is applied to the measurement
signal. There is no situation in which you would want to send this
command with
OFF
.
OFF
is only available because it is required for the
query response.
If
1|ON
is selected, it causes the error -224, “Illegal parameter value” to
occur.
Syntax
Example
CALC1:REL:AUTO ONCE
This command sets a reference
value to be used in the relative
measurement on the upper
window.
?
CALC
1
2
:MAGN
:REL
:AUTO
0|OFF
ONCE
Space
Summary of Contents for E4418B
Page 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Page 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Page 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Page 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Page 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Page 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Page 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Page 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Page 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Page 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Page 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Page 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Page 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Page 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Page 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Page 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...