![HP E4401B Programmer'S Manual Download Page 270](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548270.webp)
5-116
Chapter 5
Language Reference
SYSTem Subsystem
SYSTem Subsystem
This subsystem is used to set the controls and parameters associated
with the overall system communication. These functions are not related
to instrument performance.
HP-IB Address
:SYSTem:COMMunicate:GPIB[1]|2|3|4[:SELF]:ADDRess <integer>
:SYSTem:COMMunicate:GPIB[1]|2|3|4[:SELF]:ADDRess?
Sets and queries the HP-IB address.
NOTE
This command applies only to analyzers having Option A4H, HP-IB
and Parallel I/O).
Factory Preset
and *RST:
It is set to 18 by
:SYSTem:PRESet:PERSistent
, which
sets the persistent state values to their factory defaults.
This command is persistent. The term persistent
means that the command retains the setting previously
selected, even through a power cycle.
Range:
Integer, 0 to 30
Front Panel
Access:
System, Remote Port
Serial Port DTR Setup
:SYSTem:COMMunicate:SERial[1]|2|3|4|5|6|7|8:CONTrol:DTR
OFF|ON|IBFull
:SYSTem:COMMunicate:SERial[1]|2|3|4|5|6|7|8:CONTrol:DTR?
Sets the hardware pacing scheme. It controls the number of bytes in the
input buffer.
Off - holds the DTR line in the unasserted (off) condition
On - holds the DTR line in the asserted (on) condition
IBFull - selects the input buffer full mode for the DTR line. The
IBFull parameter sets the DTR line to indicate when the device is
ready to receive. When the number of received bytes in the input
buffer of the device reaches the stop threshold, the device will
unassert the DTR line. When the number of bytes has been reduced
to the start threshold, the device will assert DTR indicating that it
can receive input again. The device will also monitor the state of
CTS and will stop transmission if the line becomes unasserted.
Summary of Contents for E4401B
Page 6: ...vi ...
Page 14: ...xiv Contents ...
Page 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Page 65: ...3 1 3 Programming Examples ...
Page 151: ...4 1 4 Programming Command Cross References ...
Page 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Page 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Page 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Page 368: ...7 1 7 Error Messages ...
Page 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...