Chapter 5
5-97
Language Reference
SENSe Subsystem
SENSe:SWEep Subsection
Sweep Time
[:SENSe]:SWEep:TIME <time>
[:SENSe]:SWEep:TIME?
Specifies the time in which the instrument sweeps the display.
Remarks:
A span value of 0 Hz causes the analyzer to enter zero
span mode. In zero span the X-axis represents time
rather than frequency. In this mode, the sweep time
may be set to faster values when Option AYX (fast
digitized time domain sweeps) is installed.
Range:
5 ms to 2000 seconds without Option AYX, fast
digitized time domain sweeps; 20
µ
s to 2000 seconds
with Option AYX, in zero span only
Default Unit:
seconds
Front Panel
Access:
Sweep, Sweep Time Auto Man
Automatic Sweep Time
[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TIME:AUTO?
Automatically selects the fastest sweep time for the current settings.
Factory Preset
and *RST:
On
Front Panel
Access:
Sweep, Sweep Time Auto Man
Sweep Time Mode
[:SENSe]:SWEep:TIME:AUTO:MODE SRESponse|SANalyzer
[:SENSe]:SWEep:TIME:AUTO:MODE?
Specifies the type of automatic coupling for the fastest sweep time at
the current settings.
Stimulus response
Spectrum analyzer
Factory Preset
and *RST:
SANalyzer
Front Panel
Access:
Sweep, Sweep Coupling SR SA
Summary of Contents for E4401B
Page 6: ...vi ...
Page 14: ...xiv Contents ...
Page 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Page 65: ...3 1 3 Programming Examples ...
Page 151: ...4 1 4 Programming Command Cross References ...
Page 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Page 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Page 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Page 368: ...7 1 7 Error Messages ...
Page 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...