5-56
Chapter 5
Language Reference
MEASurement Subsystem
MEASurement Subsystem
MEASurement subsystem commands apply only to measurements
found in the
MEASURE
menu. These commands primarily control
measurements in single mode.
Use the single commands in this subsystem to quickly make
measurements using the factory default instrument settings.
MEASure
commands stop the present measurement and set up the
instrument for the specified measurement using the factory default
values. Other SCPI communication is blocked until the measurement is
complete. After the data is valid, the scalar result is returned for the
specified measurement.
If you need to change some of the measurement parameters from the
factory default settings, set up the measurement with the CONFigure
command. Use the commands in the SENSe: subsystem to change the
settings. Then use the
:READ?
command, or the
:INITiate
and
:FETCh?
commands to initiate the measurement and query the results.
Measure Main, Lower, and Upper Channel Power
:MEASure:ACPower?
This command returns scalar results of main channel power, lower
channel power (relative), and upper channel power (relative).
Remarks:
The main channel power is returned in the current
amplitude units, and the lower and upper channel
results are always returned in dB. The results are
returned in a comma-separated list.
Measure Main Channel Power
:MEASure:ACPower:MAIN?
This command returns the value of main channel power in the current
amplitude units.
Front Panel
Access:
MEASURE, ACP
Measure Lower Channel Power
:MEASure:ACPower:LOWer?
This command returns the value of the lower channel power relative to
the main channel power, in dB.
Summary of Contents for E4401B
Page 6: ...vi ...
Page 14: ...xiv Contents ...
Page 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Page 65: ...3 1 3 Programming Examples ...
Page 151: ...4 1 4 Programming Command Cross References ...
Page 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Page 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Page 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Page 368: ...7 1 7 Error Messages ...
Page 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...