Chapter 5
5-37
Language Reference
COUPle Subsystem
COUPle Subsystem
Some measurement settings are automatically coupled together to
optimize speed and accuracy. These commands control that coupling.
COUPle the Function to Other Settings
:COUPle ALL|NONE
The instrument can automatically couple instrument settings together
for accurate measurements and optimum dynamic range. This
command is used to override the coupling for special measurement
needs.
Factory Preset
and *RST:
All
Remarks:
The following list of analyzer functions can either be
automatically coupled, or not coupled (manually set):
Residual bandwidth
Video bandwidth
Sweep time
Center frequency step
Attenuation
Marker count resolution
Source attenuation mode
Source amplitude step
COUPle NONE
causes these functions into the manually
set (not coupled) mode.
COUPle ALL
causes the
functions into the auto coupled mode, and also causes
the sweep coupling mode into SA (couple all).
Front Panel
Access:
Auto Couple
Summary of Contents for E4401B
Page 6: ...vi ...
Page 14: ...xiv Contents ...
Page 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Page 65: ...3 1 3 Programming Examples ...
Page 151: ...4 1 4 Programming Command Cross References ...
Page 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Page 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Page 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Page 368: ...7 1 7 Error Messages ...
Page 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...