![HP E3494A Installation And Service Manual Download Page 87](http://html.mh-extra.com/html/hp/e3494a/e3494a_installation-and-service-manual_161547087.webp)
Problems with the LAN Interface
If you cannot verify LAN communication
If you cannot verify connection using the procedure in "To verify LAN
communication", or if the commands are not accepted by the processor probe:
Make sure that you have connected the HP processor probe to the
proper power source and that the power light is lit.
Make sure that the LAN cable is connected. Refer to your LAN
documentation for testing connectivity.
Make sure that only one of the LAN ports is connected.
Make sure the HP processor probe communication configuration
switches are set correctly. Unplug the HP processor probe power cord,
then plug it in again to make sure the switch settings are read correctly
by the HP processor probe.
Make sure that the HP processor probe’s IP address is set up correctly.
Use the RS-232 port to verify this that the IP address is set up correctly.
When you are connected to the RS-232 port, run performance
verification on the HP processor probe’s LAN interface with the "pv"
command.
It’s also possible for there to be a problem with the HP processor probe
firmware while the LAN interface is still up and running. In this case,
you must reboot the HP processor probe by disconnecting power to the
HP processor probe and reconnecting it again.
Solving Problems
If you cannot verify LAN communication
79
Summary of Contents for E3494A
Page 3: ...iii ...
Page 4: ...iv ...
Page 8: ...Contents viii ...
Page 9: ...Part 1 Installation Installation 1 ...
Page 10: ...Installation 2 ...
Page 11: ...1 Connecting to the Host Computer Connecting to the Host Computer 3 ...
Page 34: ...Connecting to the Host Computer To verify serial communications 26 ...
Page 35: ...2 Connecting to the Target System Connecting to the Target System 27 ...
Page 39: ...Part 2 Using the HP processor probe Using the HP processor probe 31 ...
Page 40: ...Using the HP processor probe 32 ...
Page 63: ...Part 3 Reference Reference 55 ...
Page 64: ...Reference 56 ...
Page 65: ...4 Designing a Target System Designing a Target System 57 ...
Page 72: ...Designing a Target System PowerPC 603 JTAG Interface Connections and Resistors 64 ...
Page 73: ...5 Specifications and Characteristics Specifications and Characteristics 65 ...
Page 77: ...6 Updating Firmware Updating Firmware 69 ...
Page 83: ...7 Solving Problems Solving Problems 75 ...
Page 102: ...Solving Problems To obtain replacement cables 94 ...