A returned value of 5 (calibration aborted) is given if either:
a. The command is received when the module is in voltage calibration or
resistance calibration mode, or
b. The calibration in process has entered more than 50 values without a
successful calibration.
•
If the internal calibration resistor is not calibrated, the channel is not marked as
calibrated following the completion of calibration (see the
CALibration:CONDition?
query). The new calibration constants are used, however.
•
When the channel current calibration is completed, the new calibration constants
are used by the module, but not stored in non-volatile memory.
•
If a calibration is interrupted before completion (that is, by another command
being received), no constants will be changed.
•
Refer to Appendix D, “Voltage/Current Output Adjustment”, starting on page 157,
for additional information about this command.
•
Related Commands:
CALibration:STORe
,
CALibration
n
:CONFigure:CURRent?
CALibration
n
:VALue:VOLTage?
CALibration
n
:VALue:VOLTage? <
value
>
enters a measured calibration value and
returns a value indicating the present state of the voltage calibration process.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
CALibration
n
keyword
1 to 8 or 1 to 16
1
<
value
>
numeric (float 32)
0 to 19.2 V
none
Returns: unsigned 16-bit integer (uint16).
Comments
•
In the keyword,
CALibration
n, the valid range of n is:
1 to 8 for
E1418A, E1418A Option 002
1 to 16 for
E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
•
This command is used together with the
CALibration
n
:CONFigure:VOLTage?
query
to perform voltage mode calibration. Following the
CAL
n
:CONF:VOLT?
query, this
command is used repetitively until the return value indicates that the calibration is
complete.
•
The <value> parameter is the measured output voltage. A minimum of
9 measurements must be taken to calibrate a voltage output channel
(a maximum of 50 measurements may be required).
CALibration
73 HP E1418A SCPI Command Reference
Chapter 3
Summary of Contents for E1418A
Page 6: ...Notes 6 Contents HP E1418A 8 16 Channel D A Converter Module ...
Page 10: ...Notes 10 HP E1418A User s Manual ...
Page 12: ...12 HP E1418A User s Manual ...
Page 105: ...TRIGger 105 HP E1418A SCPI Command Reference Chapter 3 ...
Page 111: ...Notes HP E1418A Command Quick Reference 111 HP E1418A SCPI Command Reference Chapter 3 ...
Page 135: ...135 HP E1418A Register Based Programming Appendix B ...
Page 157: ...Notes 156 HP E1418A Error Messages Appendix C ...
Page 170: ...Notes Appendix D Voltage Current Output Adjustment 169 ...
Page 174: ...Figure E 1 8 Channel Disassembly 172 Configuration and Disassembly Appendix E ...
Page 175: ...Figure E 2 16 Channel Disassembly Appendix E Configuration and Disassembly 173 ...
Page 192: ...Notes 192 HP E1418A 8 16 Channel D A Converter Module Index ...