
NOTE
Test failures can be caused by improper selection of the interface select
code, primary address setting, and/or secondary address setting. Verify
proper address selection before troubleshooting.
Example: Self-Test
10 OUTPUT 70903;"*TST?"
!Send the self-test command
20 ENTER 70903;A
!Enter the test result
30 PRINT A
!Display the result
40 END
Operation
Verification
Tests
There are no separate operation verification tests for the HP E1411
multimeter. Use the Performance Verification tests for post-repair
checkout.
Performance
Verification
Tests
Performance verification tests are used to check the multimeter’s electrical
performance against the specifications in Appendix A - Specifications of the
HP E1326A/E1411A User’s Manual or the E1326B/E1411B User’s Manual
as the performance standards. These tests are suitable for incoming
inspection and troubleshooting. The performance verification tests for the
HP E1411 multimeter are:
•
Test 4-1: DC Voltage Test (Zero Volt Input)
•
Test 4-2: DC Voltage Test (DCV Input)
•
Test 4-3: AC Voltage Test
•
Test 4-4: Resistance Test (4-Wire Ohms)
Chapter 4
Verification Tests 21
Summary of Contents for E1411A
Page 4: ...Notes ...
Page 8: ...Notes 8 HP E1411A B Service Manual ...
Page 10: ...10 HP E1411A E1411B Service Manual ...
Page 36: ...Notes 36 Verification Tests Chapter 4 ...
Page 48: ...Notes 48 Adjustments Chapter 5 ...
Page 51: ...Figure 6 1 HP E1411A Replaceable Parts Chapter 6 Replaceable Parts 51 ...
Page 53: ...Figure 6 2 HP E1411B Replaceable Parts Chapter 6 Replaceable Parts 53 ...
Page 56: ...56 Manual Changes Chapter 7 ...