HP D300 Digital Dispenser
54
Figure 5-28
Titration1 pane showing Modified Range titration1 layout
Alternatively, if a 100 µM maximum concentration is desired instead of the 1 µM limit described
in step
, above, the Modified Range function can be used in the Fluid Extension mode.
9.
Set the Low (µM) at 0.00002 µM if you have not, and set the High (µM) to 100 µM. Note that
the 100 µM text is red, indicating that it is outside the valid concentration range.
10.
Click the
Modified Range
button. The Valid Range Modification dialog box appears.
11.
Select the
Fluid Extension
option.
Figure 5-29
Valid Range Modification dialog box
Note that the wizard suggests using an additional fluid at a concentration of 0.1 mM, which
provides a valid concentration range of 0.00002 µM to 100 µM, exactly what is needed.
12.
Click the
Use Modification
button.
Note that the fluid table now has two entries in the Conc.(mM) column with two Dispenseheads
when this plate layout is run. This is shown in