![HP CodeMaster XL+ M1722A User Manual Download Page 81](http://html.mh-extra.com/html/hp/codemaster-xl-m1722a/codemaster-xl-m1722a_user-manual_161596081.webp)
A-1
A
Installation and Setup
1
Installation
The CodeMaster XL+ is ready for operation when the following tasks have been
properly performed:
o
Install battery.
o
Charge battery (for 24 hours).
o
Install paper.
o
Make sure that the paddle set connector is seated and locked.
o
Select configuration settings; set date and time.
Line Voltage Settings
The defibrillator automatically adjusts to the line voltage that is supplied (from 100–
230 VAC ± 15% at 50/60 Hz). No manual setting or adjustment is required.
Use only three wire power cords with three-pronged grounded
plugs. Make sure that the outlet accepts the three-pronged plugs
and is grounded. Never adapt a three- pronged plug to fit a two-
pronged outlet.
Installing and Charging the Battery
To install the battery, refer to the battery replacement procedure in “Replacing the
Battery” (page 8-5). After installing the battery, connect the power cord to the back
of the defibrillator, then plug the cord into an AC outlet. The green
and
indicators on the front panel should light up. (The
indicator lights when the instrument is plugged into AC power; the
indicator is on when the battery is installed and the instrument is plugged into AC
power.)
WARNING
AC
BATT
AC
BATT