Monitoring
Using Leads to Monitor
3-2
Preparing the Leads for Monitoring
The CodeMaster XL+ can be configured to use either a 3-wire or a 5-wire patient
cable. Use setup menu 2 as described in Appendix A to select the patient cable type
(3-wire, 5-wire).
3-Wire Patient Cable
Table 3-2, “3-Wire Electrode Placement,” on page 3-2 describes typical lead-wire
placement using the 3-wire patient cable. Table 3-3, “Lead Formation,” on page 3-3
shows how the individual leads are formed using the individual leadwires.
PADS
Pads adaptor cable:
M1750A/B
•
ECG Normal
•
Synchronized
Cardioversion
You can use these
pads for pacing, how-
ever you must select
LEADS as the moni-
tor source during pac-
ing.
M1749A
PADDLES
Standard with instrument
M1746A
Emergency ECG
Monitoring
N/A
Table 3-2
3-Wire Electrode Placement
Electrode
Placement
RA/White
Near right midclavicular line, directly below the clavicle.
LA/Black
Near the left midclavicular line, directly below the clavicle.
LL/Red
Below the left pectoral muscle on the left midclavicular line.
Table 3-1
Cardiac Monitoring Configurations
For this ECG
source...
Use this cable type...
In this monitoring
application...
Electrode P/N