Chapter 6, HP 8924E Performance Verification
CDMA Analyzer Average Power Level Accuracy Performance Test 30
N:\mkt\MANUALS\HP8924C\ALR\24e_pt.fb
CDMA Analyzer Average Power Level Accuracy
Performance Test 30
Description
The CDMA average power level accuracy is verified by comparing the measured
power in a CDMA signal with the power level measured by a power meter. The
signal is first measured by the power meter for a reference and then the same
measurements are performed using the HP 8924E’s analyzer. A dual directional
coupler is used to minimize mismatch uncertainty.
Calibration Setup
Signal Generator
OUTPUT
INPUT 50
Ω
Measuring Receiver
SENSOR
Incident
Test Port
50
Ω
Termination
OUTPUT
Sensor Module
No Connection
Direct Connection (no cables)
Dual Directional Coupler
Summary of Contents for 8924E