482
Chapter 8, Programming The Call Processing Subsystem
Programming The CALL CONTROL Screen
[] Connect
When lit, the
Connect
annunciator indicates that the mobile station is connected
to the Test Set on a voice channel.
The
Connect
annunciator is not programmable.
The state of the
Connect
annunciator is reflected in the Call Processing Status
Register Group Condition Register bit 5. See “Status Reporting” in the
Application Guide for further information.
NOTE:
When the
CALL CONTROL
screen is displayed and the Call Processing Subsystem is in the
Connect
state, the host firmware constantly monitors the mobile station’s transmitted
carrier power. If the power falls below 0.0005Watts the Test Set will terminate the call and
return to the
Active
state. The mobile station’s transmitted carrier power is monitored on
all Call Processing Subsystem screens except the
ANALOG MEAS
screen.
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...