468
Chapter 8, Programming The Call Processing Subsystem
Using the Call Processing Subsystem’s Remote User Interface
Conditioning the Test Set for Call Processing
It is recommended the control program perform the following steps when first
entering the Call Processing Subsystem (that is, the first time the
CALL CONTROL
screen is selected during a measurement session).
•
Zero the RF Power meter to ensure accurate power meter measurements.
There are two reasons for zeroing the RF power meter:
a. When any Call Processing Subsystem screen is displayed (except the ANALOG
MEAS screen) and the Call Processing Subsystem is in the
Connect
state, the host
firmware constantly monitors the mobile station’s transmitted carrier power. If the
power falls below 0.0005 Watts the error message
RF Power Loss indi-
cates loss of Voice Channel
will be displayed and the Test Set will ter-
minate the call and return to the
Active
state. Zeroing the power meter cancels
any inherent dc offsets that may be present within the power meter under zero pow-
er conditions. This ensures that the host firmware makes the correct decisions re-
garding the presence of the mobile stations’s RF carrier.
b. Zeroing the power meter establishes a 0.0000 W reference for measuring the mobile
station’s RF power at the RF IN/OUT port. This ensures the most accurate RF pow-
er measurements of the mobile stations’s RF carrier at different power levels.
Example
OUTPUT 714;"RFG:AMPL:STATE OFF"
OUTPUT 714;"DISP RFAN;:RFAN:PME:ZERO"
OUTPUT 714;"RFG:AMPL:STATE ON"
NOTE:
Ensure that no RF power is applied to the
RF IN/OUT
port when the power meter is being
zeroed. Set the
RFGenerator
amplitude to
OFF
before zeroing the power meter and then
set the
RFGenerator
amplitude to
ON
after zeroing the power meter.
•
Couple the variable frequency notch filter to AFGen1.
This step is only required if audio testing is to be done on the mobile station. This step
couples the variable frequency notch filter to the output frequency of AFGen1 (audio
frequency generator #1). The notch filter is used when making SINAD measurements.
AFGen1 is used to generate the audio tone for the SINAD measurement. Coupling the
notch filter to the audio source ensures the most accurate measurement.
Commands:
OUTPUT 714;"DISP CONF;:CONF:NOTC ’AFGEN1’"
Call Processing Subsystem HP-IB Error Messages
The Call Processing Subsystem HP-IB error messages are detailed in
through
.
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...