150 kHz to 1300 MHz

Technical Specifications

Outstanding signal
characterization

All parameters describe performance in automatic operation or
properly set manual conditions.  Specifications describe the
instrument’s warranted performance.  Supplemental characteris-
tics (shown in italics) are intended to provide information useful
in applying the instrument by giving typical, but non-warranted,
performance parameters.

RF Input

Frequency Range:

 150 kHz to 1300 MHz.

Operating Level:

150 kHz to 650 MHz: 12 mV

rms

 (

25 dBm) to 7 V

rms

(1 W

peak

).

650 MHz to 1300 MHz: 22 mV

rms

 (

20 dBm) to 7 V

rms

(1 W

peak

).

Supplemental Characteristics

Tuning: Automatic, track (frequencies >10 MHz),

manual frequency entry.

Acquisition Time (automatic operation): ~1.5

second.

Input Impedance:

 

50 

 nominal.

Maximum Safe Input Level:

AC: 35 V

rms

 (25 W for source SWR <4)

DC: 40 V

Frequency Modulation

Rates:

150 kHz to 10 MHz: 20 Hz to 10 kHz.
10 MHz to 1300 MHz: 20 Hz to 200 kHz

1

.

Deviations:

150 kHz to 10 MHz: 40 kHz peak maximum.
10 MHz to 1300 MHz: 400 kHz peak maximum

1

.

Accuracy:

2

250 kHz to 10 MHz: ±2% of reading ±1 digit, 20 Hz

to 10 kHz rates.

10 MHz to 1300 MHz: ±1% of reading ±1 digit, 50 Hz

to 100 kHz rates.

±5% of reading ±1 digit, 20 Hz to 200 kHz rates.

Demodulated Output Distortion:

3

400 kHz to 10 MHz: <0.1% THD, deviations <10 kHz.

10 MHz to 1300 MHz: <0.1% THD, rates and

deviations <100 kHz.

AM Rejection (for 50% AM at 400 Hz and

1 kHz rates):

2

 <20 Hz peak deviation measured

in a 50 Hz to 3 kHz BW.

Residual FM (50 Hz to 3 kHz BW):

 <8 Hz

rms

 at

1300 MHz, decreasing linearly with frequency
to <1 Hz

rms

 for 100 MHz and below.

Supplemental Characteristics

Maximum Deviation Resolution:

1 Hz, <4 kHz deviation.
10 Hz, 4 kHz to 40 kHz deviation.
100 Hz, 40 kHz to 400 kHz deviation.

Resolution: Is increased one digit with 750 

µ

s

de-emphasis and pre-display on.

Demodulated Output Distortion: 150 to 400 kHz;

<0.3% THD, deviations <10 kHz.

Detectors: + peak, 

-

 peak, and average (rms sine

wave calibrated).

Demodulated Output Across an Open Circuit

(600 

 output impedance):

4

1 mV/Hz when resolution is 1 Hz.
0.1 mV/Hz when resolution is 10 Hz.
0.01 mV/Hz when resolution is 100 Hz.

Stereo Separation (50 Hz to 15 kHz): >47 dB.

Phase Modulation

Carrier Frequency:

 10 MHz to 1300 MHz.

Rates:

 200 Hz to 20 kHz.

1 Maximum rate 20 kHz and peak deviation 40 kHz with 750 

µ

s

de-emphasis filter.

2 Peak residuals must be accounted for in peak readings.

3 With 750 

µ

s de-emphasis and pre-display “off,” distortion is not

specified for modulation autopilots >4 V peak. This can occur near
maximum deviation for a measurement range at rates <2 kHz.

4 For optimum flatness, cables should be terminated with their

characteristic impedance.

Summary of Contents for 8901A

Page 1: ...kHz rates Demodulated Output Distortion 3 400 kHz to 10 MHz 0 1 THD deviations 10 kHz 10 MHz to 1300 MHz 0 1 THD rates and deviations 100 kHz AM Rejection for 50 AM at 400 Hz and 1 kHz rates 2 20 Hz peak deviation measured in a 50 Hz to 3 kHz BW Residual FM 50 Hz to 3 kHz BW 8 Hzrms at 1300 MHz decreasing linearly with frequency to 1 Hzrms for 100 MHz and below Supplemental Characteristics Maximum...

Page 2: ...ld be terminated with their characteristicimpedance 5 For peak measurements only AM accuracy may be affected by distortion generated by the modulation analyzer In the worst case this can decrease accuracy by 0 1 of reading for each 0 1 of distortion 6 After30 daywarm up Standard Option 002 Aging Rate 1 x 10 6 mo 1 x 10 9 day Temperature Effects 2 x 10 7 C 2 x 10 10 C Line Voltage Effects 1 x 10 6 ...

Page 3: ...h one pole AM Output 10 kΩ impedance 4 V to 0 V into an open circuit 8 mV dc coupled 16 kHz bandwidth one pole Recorder Output DC voltage proportional to peak voltage of the modulation output 1 kΩ imped ance 0 V to 4 V for each resolution range into an open circuit IF Output 50 Ω impedance 150 kHz to 2 5 MHz 27 dBm to 3 dBm 10 MHz Reference Output 50 Ω impedance TTL levels 0 V to 2 2 V into an ope...

Page 4: ...repair this product must be returned to a service facility designated by HP Buyer shall prepay shipping charges to HP and HP shall pay shipping charges duties and taxes for products returned to HP from another country HP warrants that its software and firmware designated by HP for use with an instrument will execute its programming instructions when properly installed on that instrument HP does no...