Mixer Operating Parameters
Mixer Operating Parameters
Operate the HP 11970A mixers within the following operating parameters:
Maximum RF Input
Power
CW: No greater than +20 dBm
Pulse: No greater than +24 dBm at < 1
µ
s
Average: No greater than +20 dBm
Maximum LO Input
Power
No greater than +20 dBm
Required LO Power
The HP 11970 mixers r14 to +18 dBm at the LO input.
Connector Care
Leave the plastic waveguide flange cap on whenever the system waveguide
connectors are not connected together. This protects the mating surface from
scratches, which can degrade performance.
Proper Waveguide
Connector Alignment
If you are using Q, U, V, or W bands, waveguide flanges must be aligned
properly or measurement errors may result. Refer to Appendix B.