HP 85106D mm-Wave System
2-5
Operation
System Set Up
System Set Up
Setting up the system includes:
•
Turning on system power.
•
Setting up the system hardware state and instrument state memories.
•
Configuring the test set modules.
•
Performing an operator's check.
Turn On System
Power
1. Verify that the hardware is connected (refer to Chapter 1, “Getting
Started” and Figure 2-10 on page 2-31).
2. Turn on all system instruments power, except the HP 8510C. Wait 10
seconds, then turn on the HP 8510C.
If the HP 8510 hardware and instrument state are correct for the current
system configuration, the network analyzer begins making measurements
according to the last HP 8510C instrument state 8. Proceed to “Configuring
Port 1 and Port 2” on page 2-7".
If current HP 8510C hardware and instrument states are incorrect, the
HP 8510C may display an error message and a beeper may sound. The
HP 8510C remains in an error condition until the multiple source mode is
activated and it receives a signal it can detect. Turn the beeper off by
pressing [
SYSTEM
] {
BEEPER OFF
}. Configure and load the network analyzer
memory as described in the following procedure:
Configure Network
Analyzer Memory
The system set up procedure is greatly simplified by loading the appropriate
Machine Dump file from the mm-wave system disk supplied. The system
disk that is supplied contains files that define the specific hardware state and
instrument state for the various millimeter-wave frequency ranges.
Each Machine Dump file on this disk describes the complete HP 8510C
memory configuration including:
•
Appropriate hardware state.
•
Current and all eight instrument states.
•
Calibration sets, and calibration kits for one of the mm-wave bands or
for operation with a coaxial test set.
Summary of Contents for 85106D
Page 11: ...x HP 85106D mm Wave System ...
Page 117: ...4 24 HP 85106D mm Wave System Service and Troubleshooting Check Unratioed Power Levels ...
Page 139: ...5 22 HP 85106D mm Wave System HP 85104A Test Set Module Troubleshooting Procedures ...
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 208: ......
Page 238: ...B 2 HP 85106D mm Wave System Upgrade Paths ...
Page 243: ...HP 85106D mm Wave System v work surface installation 1 12 ...
Page 244: ... vi HP 85106D mm Wave System ...