HP 85106D mm-Wave System
1-17
Getting Started
Installing Test Set Modules
4. Gradually tighten all four screws in an X pattern to the same final
torque. This pattern insures that equal contact pressure is applied to all
four waveguide wall surfaces.
5. While tightening the screws, inspect the mating surfaces for proper
alignment by placing a lamp or white paper behind the connection. If the
flanges do not align without a gap, loosen all of the screws and start
over.
Figure 1-4
Installing Test Port Extensions