![HP 85106D Manual Download Page 236](http://html.mh-extra.com/html/hp/85106d/85106d_manual_161501236.webp)
A-2
HP 85106D mm-Wave System
HP-IB Addresses, Hardware States, and Instrument States
Table A-3
Important HP 85106D mm-Wave System Instrument States
1
System Parameter
mm-wave (waveguide)
Microwave
(coaxial)
Measurement
Unit
Q - Band
(WR-22)
U - Band
(WR-19)
V - Band
(WR-15)
W - Band
(WR-10)
Power Source #1
2
-20
-20
-25
-30
10
dBm
Power Source #2
+3
+3
+3
+3
N/A
dBm
Sweep Mode
any, except RAMP
3
or QuickStep
any
-
Z
0
1
50
Ω
Delay
waveguide
coaxial
-
Waveguide Cutoff
26.338
31.386
39.873
59.024
N/A
GHz
1.
I
nstrument states may be changed by selecting the {
FACTORY PRESET} key on the HP 8510.
2. Optimum power level for source #1 may vary from system to system. Adjust the power level to the maximum level without getting the
“IF Overload” error. (Refer to “RF Signal Power Control” in the “Operation” chapter).
3. With two synthesizers cannot be used with RAMP mode.
Table A-4
HP 85106D mm-wave System Multiple Source Operating Frequencies
Waveguide
Band
Start
Frequency
(GHz)
Stop
Frequency
(GHz)
Source #1
Source #2
Multiplier
Offset
Frequency
Multiplier
Offset
Frequency
R
1
26.5
40.0
1/2
0.000000000
1/8
0.020000000
Q
33.0
50.0
1/3
0.000000000
1/10
0.020000000
U
40.0
60.0
1/3
0.000000000
1/10
0.020000000
V
50.0
75.0
1/4
0.000000000
1/14
0.020000000
W
75.0
110.0
1/6
0.000000000
1/18
0.020000000
1. R-band can be used with the HP 83554A Source Modules and R11643A Test Set Kit,
only.
Summary of Contents for 85106D
Page 11: ...x HP 85106D mm Wave System ...
Page 117: ...4 24 HP 85106D mm Wave System Service and Troubleshooting Check Unratioed Power Levels ...
Page 139: ...5 22 HP 85106D mm Wave System HP 85104A Test Set Module Troubleshooting Procedures ...
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 208: ......
Page 238: ...B 2 HP 85106D mm Wave System Upgrade Paths ...
Page 243: ...HP 85106D mm Wave System v work surface installation 1 12 ...
Page 244: ... vi HP 85106D mm Wave System ...