![HP 85106D Manual Download Page 206](http://html.mh-extra.com/html/hp/85106d/85106d_manual_161501206.webp)
HP 85106D mm-Wave System
6-41
HP 85105A mm-Wave Controller
A2 IF Multiplexer
A2 IF Multiplexer
Procedure 12
Figure 6-24
Multiplexer Inputs
1. Switch the system to microwave operation by choosing test set address
20 for the coaxial test set.
2. Connect a BNC cable to the 20 MHz Out on the rear panel of the
HP 85102 and to the service adapter. Connect the service adapter to the
J10 TEST SET INTERCONNECT on the rear panel of the HP 85105A.
A 20 MHz signal is routed from the HP 85102 through the service
adapter and injected into the a1, a2, b1, and b2 inputs of the J10 TEST
SET INTERCONNECT.
3. Check the a1, a2, b1, and b2 user parameters. The traces should be flat
lines approximately -28 dB across the band (on the HP 8510 CRT).