HP 85106D mm-Wave System
6-39
HP 85105A mm-Wave Controller
A5 Attenuator Switch Driver Board
A5 Attenuator Switch Driver Board
Procedure 10
Figure 6-22 Attenuator Switch Driver Board
1. Remove cables W21 and W22 from the A5 attenuator switch driver
board.
2. Switch the instrument to port 1. When port 1 is active, the port 1
indicator light on the A5 board will be green and the port 2 indicator
light will be amber. The voltage measured at A5J1 is +15
±
1.5 Vdc
and the voltage measured at A5J2 is -15
±
1.5 Vdc.
3. Switch the instrument to port 2. When port 2 is active, the port 2
indicator light on the A5 board will be green and the port 1 indicator
light will be amber. The voltage measured at A5J1 is -15
±
1.5 Vdc and
the voltage measured at A5J2 is +15
±
1.5 Vdc.