6-30
HP 85106D mm-Wave System
HP 85105A mm-Wave Controller
LO Input to Leveling Amplifiers
LO Input to Leveling Amplifiers
Procedure 5
Figure 6-15 LO Input to Leveling Amplifiers
1. Run the system software. Select {
HIGHBAND
} to make the mm-wave
controller active.
2. Choose a power meter that covers the 4 to 6 GHz frequency range
(HP 436A meter and HP 8485A sensor). Calibrate the power meter and
attach a 3.5 mm (female to female) adapter to the sensor.
Summary of Contents for 85106D