HP 85106D mm-Wave System
6-29
HP 85105A mm-Wave Controller
LO Input at Rear Panel
LO Input at Rear Panel
Procedure 4
Figure 6-14 Rear Panel LO Input
1. Run the system software and select {
HIGHBAND
} to make the mm-wave
controller active.
2. Choose a power meter that covers the 4 to 6 GHz frequency range
(HP 436A meter and HP 8485A sensor). Calibrate the power meter and
attach a 3.5 mm (female to female) adapter to the sensor.
3. Disconnect the LO source cable from the back of the instrument and
attach the power sensor to the LO cable. Select {
SINGLE POINT MODE
}
from the [
STIMULUS
] menu, then using the RPG knob, slowly scan over
the frequency band. The power level should be 0 to +2 dBm.