![HP 85106D Manual Download Page 166](http://html.mh-extra.com/html/hp/85106d/85106d_manual_161501166.webp)
HP 85106D mm-Wave System
6-1
6
HP 85105A mm-Wave Controller
Introduction
The information in this chapter documents the operation, troubleshooting,
and the replaceable parts for the HP 85105A mm-wave controller.
Description of the
Instrument
The HP 85105A mm-wave controller (Figure 6-13 on page 6-28) is designed
to be used in conjunction with:
•
One HP 8510 network analyzer
•
Two Synthesized Sources
•
Two HP 85104A mm-wave test set modules
•
One coaxial test set (optional)
The HP 85105A provides an amplified LO signal to each of the mm-wave
modules to drive the harmonic mixers. An amplified RF signal is applied to
the active test set, either port 1 or port 2 for S-parameter measurements.
The incident and reflected IF signals from the test sets are returned to the
HP 85105A amplified, output to the analyzer, processed and displayed. A
leveling signal is passed from the test set through the controller to the RF
source to provide leveled power.
Tthe coaxial test set is selected by the HP 8510 software. The HP 85105A
sends a control signal to switch the RF signal to the test set and routes the IF
from the test set to the analyzer. No new connections are required when
switching between mm-wave and microwave operation. The HP 85105A is
equipped with the IF switching and RF switch control capability for routing
these signals to either the mm-wave or microwave test set.
Summary of Contents for 85106D
Page 11: ...x HP 85106D mm Wave System ...
Page 117: ...4 24 HP 85106D mm Wave System Service and Troubleshooting Check Unratioed Power Levels ...
Page 139: ...5 22 HP 85106D mm Wave System HP 85104A Test Set Module Troubleshooting Procedures ...
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 208: ......
Page 238: ...B 2 HP 85106D mm Wave System Upgrade Paths ...
Page 243: ...HP 85106D mm Wave System v work surface installation 1 12 ...
Page 244: ... vi HP 85106D mm Wave System ...