Maintenance
7RUHSODFHWKHPHWKDQHLVREXWDQHJDVSXULILHU
97
To replace the methane/isobutane gas purifier
Materials needed:
Methane/isobutane gas purifier (G1999-80410)
Front ferrule for 1/8-inch tubing (5180-4110)
Rear ferrule for 1/8-inch tubing (5180-4116)
Tubing cutter (8710-1709)
The methane/isobutane gas purifier needs to be replaced after four tanks of reagent
gas. This frequency may vary depending on purity of the gas and care taken in
uncapping and installing the gas purifier. A large leak upstream from the gas purifier
can quickly exhaust the reduced metal of its oxygen and moisture traps.
1
To install the methane/isobutane gas purifier, follow the instructions on
the label for installation and replacement intervals.
2
Disconnect the fittings on the old filter.
3
Remove the ferrules from the tubing at the outlet of the gas purifier.
Using the tubing cutter, cut off the end of the tubing with the ferrules.
4
Install the new filter.
5
Purge the new filter.
6
Cap the old filter and prepare it to be sent for regeneration.
See the instructions on the label.
C A U T I O N
Be sure not to remove the caps until you are ready to install the gas purifier. Only remove
the caps in the gas flow to prevent contamination by air.
W A R N I N G
0HWKDQHLVIODPPDEOH([WLQJXLVKDOOIODPHVLQWKHDUHDEHIRUHWXUQLQJRQJDVIORZ
Summary of Contents for 5973 MSD
Page 1: ...H Hardware Manual HP 5973 MSD PCI NCI ...
Page 6: ...6 ...
Page 7: ...1 General Information Specifications 9 Reagent gases and regulators 10 ...
Page 13: ... Hardware Reference 13 ...
Page 25: ... Chemical Ionization Theory 3RVLWLYH WKHRU 25 ...
Page 31: ... Chemical Ionization Theory 1HJDWLYH WKHRU 31 ...
Page 34: ...34 ...
Page 92: ...92 ...
Page 114: ...114 ...
Page 120: ...120 ...
Page 122: ...H Manual Part Number G1999 90001 Copyright 1998 Hewlett Packard Printed in USA 2 98 ...