![HP 54620A User'S And Service Manual Download Page 91](http://html.mh-extra.com/html/hp/54620a/54620a_users-and-service-manual_161475091.webp)
Figure 23 shows the cursors used to measure the period of an irregular
waveform on channel 4; the period can be read from the
∆
t display in the
measurement field as 28.50
µ
s.
Measuring the Period of an Irregular Waveform
Set which cursor is
controlled by the
Entry knob—press
both to move cursors
simultaneously
Choose the form of
readout of cursor
position
Turn off
cursors on
the display
(the menu
remains)
Cursor t1
Cursor t2
Figure 23
Making Analyzer Measurements
Ways to measure data
90