![HP 54620A User'S And Service Manual Download Page 185](http://html.mh-extra.com/html/hp/54620a/54620a_users-and-service-manual_161475185.webp)
To remove the front panel
1
Remove the intensity knob by pulling straight out.
2
Disconnect the keyboard ribbon cable from the system board.
3
Use a screwdriver to release retainer tab A, and your finger to release
retainer tab B. See figure 58.
Releasing front panel from deck of instrument
When tab B is released, be careful that the sheet metal tab of front-panel
ground input clears the softkey circuit board. The circuit board may be
depressed slightly with a screwdriver to avoid damage to the circuit board.
4
Rotate the front panel out until the bottom clears the rear of the
assembly, then lift the front panel to free the hooks on top.
When installing the front panel, make sure that the power switch shaft is
aligned with its mating hole in the front panel.
The front panel swings in to engage the two retainer tabs. Before attempting
to engage the retainer tabs, make sure that the six hooks on top of the front
panel are fully engaged with their mating holes in the sheet metal.
Front Panel Removal
Tab A
Tab B
Figure 58
Replaceable Parts
To remove the front panel
184
Summary of Contents for 54620A
Page 7: ...6 ...
Page 13: ...12 ...
Page 17: ...16 ...
Page 18: ...1 Getting Started ...
Page 20: ...Using the Logic Analyzer Figure 1 Getting Started 19 ...
Page 52: ...2 Making Analyzer Measurements ...
Page 121: ...120 ...
Page 122: ...3 Solving Problems ...
Page 127: ...126 ...
Page 128: ...4 Ensuring Accurate Measurements ...
Page 150: ...5 Testing Adjusting and Troubleshooting the Analyzer ...
Page 180: ...6 Replaceable Parts ...
Page 193: ...Exploded View of Logic Analyzer Figure 60 Replaceable Parts To order a replacement part 192 ...
Page 196: ...7 Performance Characteristics ...
Page 208: ...8 Messages ...
Page 217: ...216 ...
Page 239: ......
Page 241: ...240 ...