Testing Threshold Accuracy
Testing the threshold accuracy verifies the data channel threshold
accuracy specification. You should perform this test every 24 months
or 4000 hours of operation, whichever comes first.
These instructions include detailed steps for testing the threshold
settings of channels 0-7. After testing channels 0-7, connect and test
channels 8-15, then the external trigger input. Use the detailed steps
for channels 0-7, substituting channels 8-15 or external trigger for
channels 0-7 in the instructions.
Each threshold test tells you to record the voltage reading in the
Performance Test Record located at the end of this chapter. To check
if each test passed, verify that the voltage reading you record is within
the limits listed on the Performance Test Record.
Table 13
Equipment Required
Equipment
Critical Specifications
Recommended Model/Part
Digital Multimeter
0.1 mV resolution, 0.005%
accuracy
HP 34401A
Function Generator
DC offset voltage
±
6.3 V
HP 3325B Option 002
BNC-Banana Cable
HP 11001-60001
BNC Tee
HP 1250-0781
BNC Cable
HP 10503A
BNC Test Connector, 8-by-2
169
Summary of Contents for 54620A
Page 7: ...6 ...
Page 13: ...12 ...
Page 17: ...16 ...
Page 18: ...1 Getting Started ...
Page 20: ...Using the Logic Analyzer Figure 1 Getting Started 19 ...
Page 52: ...2 Making Analyzer Measurements ...
Page 121: ...120 ...
Page 122: ...3 Solving Problems ...
Page 127: ...126 ...
Page 128: ...4 Ensuring Accurate Measurements ...
Page 150: ...5 Testing Adjusting and Troubleshooting the Analyzer ...
Page 180: ...6 Replaceable Parts ...
Page 193: ...Exploded View of Logic Analyzer Figure 60 Replaceable Parts To order a replacement part 192 ...
Page 196: ...7 Performance Characteristics ...
Page 208: ...8 Messages ...
Page 217: ...216 ...
Page 239: ......
Page 241: ...240 ...