Troubleshooting the Logic Analyzer
The service policy for this instrument is replacement of defective
assemblies. The following procedures can help isolate problems to the
defective assembly.
W A R N I N G
The maintenance described in this section is performed with power supplied
to the logic analyzer and with the protective covers removed. Only trained
service personnel who are aware of the hazards involved should perform the
maintenance. Whenever possible, perform the procedures with the power
cord removed from the logic analyzer. Read the safety summary at the back
of this book before proceeding.
C A U T I O N
Do not disconnect any cables or remove any assemblies with the power
applied to the logic analyzer, or damage to the logic analyzer can occur.
The following equipment is needed to troubleshoot the logic analyzer.
Table 11
Equipment Required
Equipment
Critical specifications
Recommended model/part
Digital multimeter
1 mV resolution, accuracy
±
0.05%
HP 34401A
Oscilloscope
100 MHz
HP 54600
Probe
10:1 division ratio
HP 10071A
Dummy load
Compatible with power
supply
HP 54600-66504
164
Summary of Contents for 54620A
Page 7: ...6 ...
Page 13: ...12 ...
Page 17: ...16 ...
Page 18: ...1 Getting Started ...
Page 20: ...Using the Logic Analyzer Figure 1 Getting Started 19 ...
Page 52: ...2 Making Analyzer Measurements ...
Page 121: ...120 ...
Page 122: ...3 Solving Problems ...
Page 127: ...126 ...
Page 128: ...4 Ensuring Accurate Measurements ...
Page 150: ...5 Testing Adjusting and Troubleshooting the Analyzer ...
Page 180: ...6 Replaceable Parts ...
Page 193: ...Exploded View of Logic Analyzer Figure 60 Replaceable Parts To order a replacement part 192 ...
Page 196: ...7 Performance Characteristics ...
Page 208: ...8 Messages ...
Page 217: ...216 ...
Page 239: ......
Page 241: ...240 ...