Chapter 3 Programming Your Counter for Remote Operation
Status Reporting
Programming Guide
3-35
3
Transition filters are unaffected by *CLS or queries. Transition filters are
set to default values by :STATus:PRESet and power-on.
To write to the Operation Status transition filter registers use:
:STATus:OPERation:PTRansition
:STATus:OPERation:NTRansition
To read these registers use:
:STATus:OPERation:PTRansition?
:STATus:OPERation:NTRansition?
Event Register
An event register captures changes in conditions.
An event register bit (event bit) is set TRUE when an associated event
occurs. These bits, once set, are “sticky.” That is, they cannot be cleared
even if they do not reflect the current status of a related condition, until
they are read.
Table 3-6. Transition Filter Definition
Positive
Transition
Filter Bit
Negative
Transition
Filter Bit
Transition Which Causes the Event-Bit to be set TRUE
TRUE
FALSE
positive transition
FALSE
TRUE
negative transition
TRUE
TRUE
either a positive or negative transition
FALSE
FALSE
neither transition (event reporting is disabled)
Summary of Contents for 53150A Series
Page 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Page 2: ......
Page 10: ...Contents x Programming Guide ...
Page 11: ...1 Before You Start ...
Page 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Page 21: ...2 Command Summary A Quick Reference ...
Page 41: ...3 Programming Your Counter for Remote Operation ...
Page 101: ...4 Command Reference A Dictionary ...
Page 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Page 163: ...5 Errors ...
Page 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Page 184: ......
Page 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...