Chapter 4 Command Reference
[:SENSe]:FUNCtion Subtree
Programming Guide
4-27
4
[:SENSe]:FUNCtion Subtree
This subtree controls the sensor functions.
COMMAND
[:SENSe]:FUNCtion[:OFF] . . . <sensor_function>[,<sensor_function>]
Sets or queries the sensor functions to be sensed by the Counter.
The <sensor_function> strings are:
"[XNONe:]FREQuency [1 | 2]"
"[XNONe:]POWer [2]"
QUERY RESPONSE
•
The query form of this command returns a comma-separated list of
functions that are OFF.
•
The string omits default nodes (XNONe) and uses short-form
mnemonics. If the channel specifier(s) are set to default value(s), no
channel specifier is returned in response. If the channel specifier(s)
are not set to default value(s), they are returned in the response with
a single space separating the first channel specifier from the function
name.
For example:
–
“FREQ” is returned for frequency on Channel 1.
–
“FREQ 2” is returned for frequency on Channel 2.
COMMENTS
•
*RST: “FREQ 2”
•
This command can be used to turn individual function(s) OFF without
affecting other functions.
•
If the optional channel specification is omitted from the
<sensor_function>, a default channel selection is made. For Frequency
and Power, the default is Channel 2.
RELATED
FRONT-PANEL
KEYS
Various
Summary of Contents for 53150A Series
Page 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Page 2: ......
Page 10: ...Contents x Programming Guide ...
Page 11: ...1 Before You Start ...
Page 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Page 21: ...2 Command Summary A Quick Reference ...
Page 41: ...3 Programming Your Counter for Remote Operation ...
Page 101: ...4 Command Reference A Dictionary ...
Page 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Page 163: ...5 Errors ...
Page 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Page 184: ......
Page 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...