Chapter 4 Command Reference
[:SENSe] Subsystem
4-26
Programming Guide
4
COMMAND
[:SENSe]:FREQuency:RESolution . . . <numeric_value>[<frequency_unit>]
Sets the resolution of the frequency measurement.
The allowable settings for
<numeric_value>
and
<frequency unit> are:
1 Hz, 10 Hz, 100 Hz, 1 KHz, 10 KHz, 100 KHz, and 1 MHz.
COMMENTS
•
*RST: 1 Hz
•
This command does not affect the unit multipliers of any
measurement queries.
•
If no unit is specified, the global frequency unit in effect is used.
RELATED
FRONT-PANEL
KEYS
Resol
COMMAND
[:SENSe]:FREQuency:TRACking . . . <character_program_data>
Selects one of three signal-tracking modes (SLOW | FAST | OFF). When
TRACking is set to SLOW, the Counter applies two tracking routines after
each measurement; when it is set to FAST, it uses only one tracking
routine; when it is set to OFF, it does not use either of the tracking
routines.
COMMENTS
•
*RST: FAST
•
The SLOW setting provides the most accurate signal tracking but
yields the smallest number of measurements in a given period of time.
•
The FAST setting centers the IF in the IF bandwidth after each
measurement to improve tracking.
RELATED
FRONT-PANEL
KEYS
None
Summary of Contents for 53150A Series
Page 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Page 2: ......
Page 10: ...Contents x Programming Guide ...
Page 11: ...1 Before You Start ...
Page 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Page 21: ...2 Command Summary A Quick Reference ...
Page 41: ...3 Programming Your Counter for Remote Operation ...
Page 101: ...4 Command Reference A Dictionary ...
Page 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Page 163: ...5 Errors ...
Page 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Page 184: ......
Page 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...