Chapter 4 Command Reference
:MEASure Subsystem
4-12
Programming Guide
4
COMMAND
:CONFigure[:SCALar]:<function> <parameters> [,<source_list>]
Configures the instrument to perform the specified function but does not
initiate the measurement.
COMMENTS
•
Use :INITiate:FETCh? or :READ? to make and query a measurement.
•
Parameters (other than <source_list>) can be defaulted by
substituting the keyword DEFault. The <source_list> parameter can
be defaulted by omitting it. The default values are specified by the
particular function description.
•
This command defaults several Counter settings. To change the
function only, while leaving all other Counter settings as they are, use
[:SENS]:FUNC[:ON].
•
If an <expected_value> parameter is outside the measurement
capabilities of the Counter model, an error is generated, and the
command does not execute.
•
For :POWer, <resolution> must be defaulted or set to 0.01[dB].
•
See “Descriptions of the Measurement Functions” on page 4-16 for a
description of each of the measurement functions.
•
See Table 4-2 for a summary of the <function>, <parameters>, and
<source_list> for each of the measurement functions.
COMMAND
:CONFigure?
Queries the function configured by the last :CONFigure command or
:MEASure query.
QUERY RESPONSE
•
A string of the form: "<function> <parameters>[,<source_list>]." The
leading colon is omitted from the <function>.
•
The response is unaffected by *RST, recall, and [:SENS]:FUNC.
COMMENTS
•
If the instrument state has changed through commands other than
:CONFigure or the :MEASure query, the instrument does not track
these changes, and the query response does not reflect these changes.
•
For :POWer, <resolution> must be defaulted or set to 0.01[dB].
Summary of Contents for 53150A Series
Page 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Page 2: ......
Page 10: ...Contents x Programming Guide ...
Page 11: ...1 Before You Start ...
Page 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Page 21: ...2 Command Summary A Quick Reference ...
Page 41: ...3 Programming Your Counter for Remote Operation ...
Page 101: ...4 Command Reference A Dictionary ...
Page 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Page 163: ...5 Errors ...
Page 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Page 184: ......
Page 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...