Chapter 4 Command Reference
:INPut Subsystem
Programming Guide
4-9
4
COMMAND
:INITiate[:IMMediate]
This event command causes the instrument to initiate either a single
measurement or a block of measurements.
COMMENT
This command is an overlapped command (see IEEE 488.2, Section
12). Beginning a measurement or block of measurements with an
:INITiate[:IMMediate] sets the Pending Operation Flag to true.
Completing the measurement or block of measurements (normally or
by aborting) sets Pending Operation Flag to false.
RELATED
FRONT-PANEL
KEYS
None
:INPut Subsystem
This subsystem controls the characteristics of the Counter’s input ports.
COMMAND
:INPut:FILTer:[:LPASs][:STATe] . . . <Boolean>
Sets or queries the state of the Channel 1 low-pass filter.
QUERY RESPONSE
•
Single ASCII-encoded byte, 0 or 1.
•
A value of 0 indicates OFF; a value of 1 indicates ON.
COMMENT
*RST: OFF
RELATED
FRONT-PANEL
KEYS
Shift
+
Menu (CH1 LPF > OFF | ON)
Summary of Contents for 53150A Series
Page 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Page 2: ......
Page 10: ...Contents x Programming Guide ...
Page 11: ...1 Before You Start ...
Page 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Page 21: ...2 Command Summary A Quick Reference ...
Page 41: ...3 Programming Your Counter for Remote Operation ...
Page 101: ...4 Command Reference A Dictionary ...
Page 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Page 163: ...5 Errors ...
Page 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Page 184: ......
Page 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...