Chapter 1 Performance Tests
Test 5: Option 015/030/050/124 Channel 2 Frequency Accuracy (HP
53181A Only)
Assembly-Level Service Guide
1-65
1
3
Record the actual reading in the Performance Test Record
(Test 5, Line 1).
4
Change the synthesized signal generator’s frequency to 1.5 GHz.
Verify the Counter reads 1.5 GHz ±1 Hz
5
Record the actual reading in the Performance Test (Test 5, Line 2).
6
Disconnect the test setup.
Minimum to Maximum Accuracy Test for Options
030/050/124
Equipment
HP 8663A Synthesized Signal Generator (not needed for Option 050)
HP 8340B Sweep Oscillator
HP 1250-0780 Type N-to-BNC Connector (not needed for Option 050)
HP 10503A BNC Cable (2)
HP 10100C 50
Ω
Feedthrough
HP 11500D N-N Cable (Option 050 only)
HP 1250-1250 N(M) to SMA(F)
HP 1250-0777 N(F) to N(F)
Counter Setup
1
Cycle the POWER key to preset the Counter.
2
Set the Counter as follows:
MEASURE/Freq & Ratio,
FREQUENCY 2
Gate & ExtArm,
TIME: TIME 1.000 s
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...