take longer to return to normal. If overdelivery occurs
during administration, observe the patient closely for
signs of spinal cord compression (disorientation,
headache, transient neuralgias) and drug overdose.
8.4
Battery Operation
The battery may not be fully charged upon receipt.
Connect the Plum XL to AC power for at least eight hours
with the control dial in the OFF CHARGE setting.
If the LOW BATTERY alarm sounds, connect to AC power
immediately.
8.5
Sets and Accessories
Only PlumSets administration sets can be used with this
device.
Sets should be changed in accordance with current,
recognized guidelines of IV therapy. Discard sets per
hospital procedures.
LifeCare IV infusion sets with integral nonblood filters
are not for use in the administration of blood, blood
products, emulsions, suspensions, or any medications
not totally soluble in the solution being administered.
These medications may be administered through the
lower Y-injection site, below the filter.
When infusing at low delivery rates (5.0 mL/hr or less)
the use of thick-walled microbore Plumsets is
recommended. This will reduce the amount of the fluid
bolus that may be delivered when a distal line occlusion
is released.
Use the syringe adapter when using syringes 10 cc or
smaller on the secondary line (syringes must be larger
than 3 cc).
CAUTION: Do not operate the Plum XL with the battery
removed. The use of a properly maintained and charged
battery ensures proper operation. As always in the
event of an AC power interruption or failure verif
infusion pump settings.