Chapter 2. Basic Operation
Priming Pump Sets WITH a Drip Chamber
1
Push down on the flow stop
to
close the cassette.
2
Insert the piercing pin into the
container. Suspend the container
on an I.V. Pole.
3
Squeeze the drip chamber until
filled halfway.
4
Push down on the flow stop
to
open (priming position) the
cassette.
5
Ensure the CAIR (roller) clamp is
open prior to inserting cassette.
6
Invert the cassette approximately
45°.
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not
properly installed in the pump.
When using a vented chamber, ALWAYS use an
air-eliminating filter OR set the air sensitivity to
either ON or 2 mL.