Chapter 12. Troubleshooting
144
Hospira GemStar System Operating Manual
PROX. OCCLUSION
N
A/V
The pump has
detected sustained
negative pressure on
its proximal side.
1
Press [
SILENCE
] to quiet
the alarm for 1 minute.
2
Press [
STOP
] to place the
pump in the stop mode.
WARNING: When a
Proximal Occlusion alarm
activates, DO NOT
power-off the pump.
3
Check for sources of
proximal occlusion:
• Kinked tubing above pump
• Blockage in the container
and in the spike
• Closed clamps above
pump
• Empty container
• High torque syringe
• Other obstructions
4
Remove obstructions.
5
Press [
START
] to begin
delivery.
6
If the problem continues,
ensure slide clamp or lower
CAIR clamp is open.
Power-off the pump.
Reinstall the cassette. Turn
on the pump.
Note: Ensure the tubing is
behind the tubing guides (at the
distal end).
CAUTION: Ensure all
four cassette latches
are clearly visible after
the cassette is installed.
Refer to
Alarm Display
Message
Is the
pump
infusing?
Alarm Cause/
Condition
Remedy
Y = Yes
N = No
A = Audible alarm
V = Visual alarm
Summary of Contents for GemStar Series
Page 9: ...Notes viii Hospira GemStar System Operating Manual ...
Page 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Page 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Page 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Page 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Page 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Page 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Page 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Page 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Page 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Page 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Page 221: ...Notes 212 Hospira GemStar System Operating Manual ...