Hospira GemStar System Operating Manual
107
Chapter 9. Variable Time
GemStar
Variable Time Programming Worksheet
GIVE:
Parameter
Selections and/or Values to be Entered
Programmable
Ranges
Select Phase Type
Percentages
Rates
Doses
0.1 - 100%*
0.1 - 400 mL/hr
0.1 - 9600 mL
Unit of Measure
Concentration: mg/mL
Concentration: mcg/mL
mL
0.1 - 1000 mg/mL
1 - 1000 mcg/mL
Current Time
Confirm or correct current time.
12:00 AM - 11:59 PM
or 00:00 - 23:59
Base Rate
No
Yes Start time:
Stop time:
Rate:
12:00 AM - 11:59 PM
or 00:00 - 23:59
0.1 - 400 mL/hr
Total Daily Amount for All Phases:
(required only when programming in percentages)
———- SEE ATTACHED WORKSHEET———-
Bolus Dose
0.1 mL - 25.0 mL
0.1 mg - 2500 mg
1 mcg - 25000 mcg
Bolus Lockout
minutes
5 - 999 min
KVO
(Keep Vein Open)
No
KVO Rate: 0.0
Yes KVO
Rate:
0.0 - 5.0 mL/hr
Container Size
(Total Daily Amount X
Number of Days of
Therapy)
0.1 - 99999 mg
1 - 9999000 mcg
0.1 - 9999 mL
Air Sensitivity**
On
Alarms at approx. 0.5 mL of air
2 mL Alarms at approx. 2 mL of air
Off
No alarm
N/A
*
Sum of all phases must equal 100% when programmed in percentages.
** Air Sensitivity may not show in all setting configurations.
Summary of Contents for GemStar Series
Page 9: ...Notes viii Hospira GemStar System Operating Manual ...
Page 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Page 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Page 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Page 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Page 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Page 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Page 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Page 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Page 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Page 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Page 221: ...Notes 212 Hospira GemStar System Operating Manual ...