Hospira GemStar System Operating Manual
101
Chapter 8. mL/hr
Canceling a Piggyback Delivery
1
Press [
CHANGE
] and select
PIGGYBACKING
from the Change Menu.
2
Press [
CHANGE
] to reset the dose amount to zero.
3
Press [
YES/ENTER
] to cancel piggyback delivery.
Changing a mL/hr Only Program
To change a program:
1
From the
STOP
mode, press [
CHANGE
] and select
CHANGE PROGRAM
.
2
As each message displays, make the desired changes or press [
YES/ENTER
] to
accept the current setting.
3
Press [
START
] to begin delivery.
Can be changed at any time
•
Delivery rate
•
Container size
•
Piggybacking
•
VTBI
d
Note: You cannot change a program when the pump is in
FULL
or
CONTAINER
Lock.
d
Note: A program review is not required and is available only through the Change or
Options Menu.
Summary of Contents for GemStar Series
Page 9: ...Notes viii Hospira GemStar System Operating Manual ...
Page 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Page 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Page 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Page 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Page 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Page 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Page 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Page 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Page 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Page 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Page 221: ...Notes 212 Hospira GemStar System Operating Manual ...