• Keep the cassette door securely closed while the infuser is
not in use, to avoid cassette door damage.
• Values beyond a fields maximum hard limit will be diplayed
as dashes (-- -- --). The user must clear these fields using the
[CLEAR] key prior to entering new values.
• The Plum A+3 is to be used with an IV pole with a 6-wheel
base and a shelf.
C
LEANING
For more information on cleaning the infuser, see Section 8.
• To avoid mechanical or electronic damage, do not immerse
the Plum A+ in any fluids or cleaning solutions.
• Do not spray cleaning solutions toward any opening in the
instrument.
• Certain cleaning and sanitizing solutions may slowly degrade
components made from some plastic materials. Using
abrasive cleaners or cleaning solutions not recommended by
Hospira may result in product damage. Do not use
compounds containing combinations of isopropyl alcohol and
dimethyl benzyl ammonium chloride.
• Never use sharp objects such as fingernails, paper clips, or
needles to clean any part of the infuser.
• Do not sterilize by heat, steam, ethylene oxide (ETO), or
radiation.
• To avoid infuser damage, cleaning solutions should only be
used as directed. The disinfecting properties of cleaning
solutions vary; consult the manufacturer for specific
information.
BOLUS
RELATED
Use the following procedure to avoid the administration of a
bolus following a distal occlusion (i.e., a closed distal clamp):
• If a secondary container is in use, clamp proximal tubing
before opening cassette door.
• Open cassette door and remove the cassette.