5 FUNCTION
85
5.5
Water Sampling
This function stores sample water when a measurement malfunction occurs. During the set
water sampling time, sample water is obtained from the abnormal water sampling inlet with a
flow rate of 80 mL/min. It is necessary to connect a container to obtain sample water from the
inlet.
Only the first malfunction is obtained, and the time is recorded. Once sample water has been
obtained, the water sampling function is locked until reset. Sample water will not be obtained
even if another malfunction occurs.
The following settings and functions are available for this feature.
Displaying the SAMPLING SETTING screen
1.
Press [SAMPLING] on the SETTING 1/2 screen.
The SAMPLING screen is displayed.
The example of connection of the container to obtain a water sampling
Setting
Explanation
Initial Setting
(
Set the timing to obtain the sampling.
OFF
(
Set the time to obtain the sampling.
10 minutes
(
Obtains sample water once and releases the lock.
When a reset is performed, "LAST SAMPLING TIME"
is cleared from the display.
-
Displaying the maintenance item screen" (page 73).
MEAS. screen
MAINTENANCE screen
SETTING1/2 screen
[SETTING]
[MAINTE.]
[SAMPLING]
ERROR SAMPLE
OUTPUT
Tube
Container for sample water
Summary of Contents for TW-100
Page 1: ...Water Distribution Monitor TW 100 Instruction Manual CODE GZ0000080187H ...