7 Maintenance
82
7
Maintenance
Calibration and periodical maintenance are required in order to maintain the normal operation
and performance of the unit.
Regarding calibration, refer to "4 Calibration" (page 15).
7.1
Contact for Maintenance
7.2
Maintenance Item
The maintenance cycle in the table is the typical interval when using under standard
conditions, and does not guarantee the measurement values. They may vary according to the
circumstances or usage conditions set the proper cycle after using for a certain period of time.
Manufacturer:
HORIBA Advanced Techno Co., Ltd.
31, Miyanonishi-cho, Kisshoin Minami-ku, Kyoto 601-8306, Japan
Maintenance item
Maintenance cycle
Time to clean/replace
Subject
Description
Weekly Monthly
Every
half
year
Yearly
Every
3 years
Accord
ingly
Analyzer
unit
Cleaning of measuring cell
When dirt can be seen.
Inspection and replacement
of wiper blade rubber
When dirt can be seen.
Replacement of desiccant
agent and seal washers:
upper case
Periodical replacement
(every half year)
Replacement of desiccant
agent in measuring cell
Periodical replacement
(every half year)
Replacement of inner parts
When the measurement
value is abnormal
Replacement of UV lamp
Periodical replacement
(yearly)
When the measurement
value is abnormal
When the UV lamp
does not illuminate
Replacement of motor
Periodical replacement
(yearly)
Replacement of seal
washers: cell stay
Periodical replacement
(yearly)
Replacement of V-rings
Periodical replacement
(yearly)
Replacement of gear
Periodical replacement
(yearly)
Replacement of cell
When dirt can be seen
When it is damaged
Replacement of case
packing
Periodical replacement
(yearly)
Control
unit
Replacement of battery for
clock
Periodical replacement
(every 3 years)
Replacement of substrate
and electric parts
When malfunctioning