3 Operation
13
3.2
Starting Operation
1.
Turn the POWER switch ON.
The MEAS. screen appears after a while.
Note
When the power is turned ON, the motor for analyzer unit wipers starts operation at low speed. Some
noises and vibrations may be generated when the motor starts. Note that this is normal, and does not
mean any malfunction of the unit.
"WARM UP!" continues to be indicated until the measured values become stable (for
about 60 minutes).
2.
Set the clock.
Reference
"5.4.5 Time Adjustment" (page 43)
3.
Change the settings as necessary.
Reference
4.
Calibrate the unit.
Note
Calibration should be performed after the measured value becomes stable. Before the calibration,
perform the running-in operation for an hour or more using the sample water.
Reference
Complete the procedure for the preparation of operation. The measurement starts.
5.
Turn OFF the maintenance switch.
“WARM UP!” is indicated.
WARNING
Do not touch the cell or directly view the UV lamp of the analyzer unit when you check the
unit operation.