7 Maintenance
89
7.3.4
Replacement of Desiccant Agent in the Measuring Cell
Here describes the replacement of desiccant agent in the measuring cell of analyzer unit.
Replace the desiccant agent every 6 months.
1.
Turn ON the maintenance switch.
2.
Turn OFF the power switch.
3.
Remove the analyzer unit from the measuring tank.
4.
Lightly wipe off moisture adhered to the measuring cell with soft cloth.
5.
Remove the 2 brackets from the analyzer unit.
The following description is the replacement method for the desiccant agent of the light
source cell.
Replace the desiccant agent on the detector side by following the same procedure.
6.
Remove the screw.
7.
Remove the cell plate D.
8.
Remove the cell packing by using tweezers.
9.
Take the cell plate C and remove the desiccant agent.
Bracket
Screw
Cell plate D
Cell packing
Tweezers
Cell plate C
Desiccant agent