Dissolved oxygen measurement
19
• Setting the industrial standard (default: ISO)
The instrument can be adaptable to the industrial standards of ISO17289 (JIS K0102),
JIS K0101, JIS K0102 before 2015, and JIS K0803. When evaluating with ISO17289
(JIS K0102), select "ISO". When evaluation with JIS K0101, JIS K0102 before 2015, and
JIS K0803 select "Pre". Saturated DO levels in water at various temperature of each
standard is shown the section of technical note.
1. Press the
key to enter the setting mode.
2. Press the
keys to select the "CAL-TYPE" (standard setting) and then press
the
key.
3. To evaluate with ISO17289 (JIS K0102), select "ISO". To evaluate with JIS K0101,
JIS K0102 before 2015, and JIS K0803 select "Pre".
4. Press the
key to reflect the change.
key
key