Maintenance
93
3.
Pull out the latch knobs (2 places).
4.
Disconnect the extractor joint and motor connector, and remove the extractor.
5.
Loosen the fixing screw and remove the extraction tank.
Fig. 105 Extractor
6.
Use pure water or clean solvent S-316 to wash the extraction tank.
7.
Remove droplet inside the extraction tank with tissue paper.
Fig. 103 Removing the sample inlet
Fig. 104 Latch knobs